Preliminary Vacuum-System Estimator
Pumping speed interpolated between the six manufacturer-published points. Target-pressure estimates run from 280 to 950 mbar, using manufacturer-published pumping-speed data covering 280–1000 mbar.
Preliminary teaching tool, not sizing software. Speeds between the published points are linearly interpolated; below 280 mbar the manufacturer publishes no speed data, so no values are estimated there. Times integrate the throughput balance for a closed, leak-tight, isothermal system using the linearly interpolated manufacturer-published speed data, t = V ∫ dp / [S(p) · p], starting from the 1000 mbar end of the published table rather than from standard atmospheric pressure. Real systems are slower: hose and trap conductance, leaks, outgassing and above all evaporating solvent vapor add load. Confirm pump selection against your own process and the manufacturer’s documentation.