Impact strength from absorbed energy

Enter the corrected absorbed energy you read from the tester and the dimensions of the bar. The calculator only does the arithmetic of a general method. It is not a product claim, it does not choose a test method, and it gives no pass or fail result.

Your numbers

J
mm
mm

Result

–

Arithmetic only. The selected method determines whether the specimen outcome is reportable. This calculation does not establish test validity, method equivalence or pendulum suitability.

Display rounding does not establish measurement precision or test validity.

Impact strength a = E / (h × bN) × 1,000, with E in J and h and bN in mm, gives kJ/m².

Cross-section under the notch– Energy divided by h (J/m)– Absorbed energy as a share of the pendulum energy–

Caveat. General method arithmetic on the numbers you enter, not a statement about any material or product. The shaded band between the two marker lines is the 10% to 80% window of pendulum energy stated in the 1993 editions of the ISO Charpy and Izod methods; the current edition of your own method decides the rule. J/m is E divided by the dimension along the notch: check which dimension your method divides by before comparing it with kJ/m². Listed pendulum energies are the values on the Polytevo Series Charpy and Izod tester pages (page snapshot 15 November 2026); supplied configurations are confirmed on the quotation. Nothing you enter is stored or sent.

ACS Material LLC. Illustrative arithmetic tool.