Filter Press AFP16
Product Detail
AFP16 is a lab filter press designed for slurry-pressing simulation tests: a pneumatic cylinder reproduces plate-and-frame filter pressing with precisely controlled parameters — multi-stage pressure settings and pressing duration — for a repeatable, standardized test environment. An electronic balance under the test cylinder weighs the filtrate in real time, logging the data with on-screen parameter setting and visual display.
Key features
- Two-stage pneumatic pressingPrimary and secondary pressing stages each control 0–1.6 MPa of pressure over 0–30 min, reproducing plate-and-frame filter pressing.
- Repeatable, standardized testsMulti-stage pressure settings and pressing durations are set precisely, giving a repeatable, standardized environment for slurry-pressing simulation.
- Real-time filtrate weighingAn electronic balance (3 kg range, 0.01 g accuracy, data transfer) under the test cylinder measures the filtrate in real time.
- Curves and data loggingThe control software records the real-time total filtrate and filtrate–time curves, with parameter setting and visual data display.
- 785 mL test cylinderThe test cylinder holds 785 mL at 100 mm inner diameter × 100 mm height; the tower measures 367 × 223 × 710 mm at 38 kg.
Product specifications
| Parameter | AFP16 |
|---|---|
| Model # | AFP16 |
| SKU # | EIFP16 |
| Dimensions | 367 × 223 × 710 mm |
| Primary Pressing Pressure | 0 ~ 1.6 MPa |
| Primary Pressing Time | 0 ~ 30 min |
| Secondary Pressing Pressure | 0 ~ 1.6 MPa |
| Secondary Pressing Time | 0 ~ 30 min |
| Electronic Balance | 3 kg range, 0.01 g accuracy, data transfer |
| Cylinder Volume | 785 mL |
| Inner Diameter | 100 mm |
| Height | 100 mm |
| Control Software | Real-time total filtrate; filtrate–time curves |
| Net Weight | 38 kg |
Basic configuration
| Included with every press | Qty |
|---|---|
| Control software | 1 |
| Main unit (filter press) | 1 |
| Pressure controller | 1 |
| Electronic balance (model not fixed) | 1 |
| Air compressor (model not fixed) | 1 |
| Connection cables & accessories | — |
The electronic balance and air compressor models are not fixed and are specified per order; final configuration is confirmed at quotation.
FAQ
What does it simulate?
The AFP16 simulates slurry pressing: a pneumatic cylinder reproduces plate-and-frame filter pressing under precisely controlled pressure and duration, for repeatable, standardized filtration tests.
What can I control?
Primary and secondary pressing stages are set independently — each 0–1.6 MPa with multi-stage pressure settings and 0–30 min durations.
How is the filtrate measured?
An electronic balance under the test cylinder (3 kg range, 0.01 g accuracy, data transfer) weighs the filtrate in real time; the software records total filtrate and filtrate–time curves.
What’s included?
Each press ships with the control software, main unit, and a pressure controller, plus an electronic balance and air compressor whose models are not fixed and are specified per order, and connection cables and accessories.
How do I order?
The AFP16 is supplied by quotation: request via this page, and our engineers confirm the configuration for your slurry and reply with pricing and lead time.
Related
- Corona Polarization Device ACP30000 — 0–30 kV corona poling for polymer films.
- Muffle Furnace AMFH1200 — electric muffle furnace to 1200 °C for sintering and annealing.
- High-Low Temperature Test Chamber ABCH180 — −120 to 180 °C chamber for variable-temperature testing.
Disclaimer: ACS Material, LLC believes the information on this page is accurate and represents the best and most current information available to us. Specifications reflect the manufacturer's current datasheet; idealized values are benchmarks and final configurations are confirmed at quotation — always refer to the model datasheet for guaranteed ratings. ACS Material makes no representations or warranties, express or implied, regarding suitability for any purpose, and will not be responsible for damages resulting from use of or reliance upon this information.