ALab™ DPCD10 Portable pH/Conductivity/Dissolved Oxygen Meter
Product Detail
ALab™ DPCD10 — a portable pH, conductivity and dissolved oxygen meter with a 3.5-inch TFT color display, a new UI interface and data storage for up to 3000 groups.

The DPCD10 is a portable meter for pH/mV (ORP), conductivity (TDS, salinity and resistivity) and dissolved oxygen, with temperature measurement. Its 3.5-inch TFT color display and the new UI interface make it easy to read and to operate. Intelligent functions include an auto. timing datalogger, parameter setting and data hold; data storage is up to 3000 groups. In conductivity mode the meter supports probe constant calibration and conductivity standard solution calibration. The probes connect through a BNC socket (pH/ORP probe), a 4-pin socket (temperature/cond probe) and an 8-pin socket (DO probe), protected by silicone caps. The meter has a rechargeable lithium battery, a Type-C charging socket and protection class IP57.
Why choose the DPCD10
Specifications General — DPCD10
| Parameter | Specification |
|---|---|
| Brand | ALab™ by ACS Material |
| Model | DPCD10 |
| SKU # | ALDP0100 |
| Product Type | Portable pH / conductivity / dissolved oxygen meter |
| Display | 3.5-inch TFT color display |
| Interface | New UI interface, easy operation |
| Functions | Auto. timing datalogger, parameter setting, data holder |
| Data storage | Up to 3000 groups |
| Cond mode | Probe constant calibration; conductivity standard solution calibration |
| Protection class | IP57 |
Technical specifications DPCD10
Specifications — DPCD10
| Parameter | Specification |
|---|---|
| pH: Measurement range | (-2.000~20.000)pH |
| pH: Resolution | 0.1/0.01/0.001 |
| pH: Accuracy | ±0.002pH ±one digit |
| pH: Calibration points | 1, 2, -3pts automatic |
| pH: ATC range | (0.0~100.0)°C/(32.0~212.0)°F |
| ORP: Measurement range | (-1999.9~+1999.9)mV |
| ORP: Resolution | 0.1/1 |
| ORP: Accuracy | ±0.03%F.S ±one digit |
| Conductivity: Measurement range | 0~2000.0mS/cm, Auto-ranging: 0~19.999μS/cm; 20.0~199.9μS/cm; 200~1999μS/cm; 2.00~19.99mS/cm; 20.00~199.9mS/cm; 200.0~2000mS/cm (K=10). Match test range: 0~200mS/cm. TDS: 0~100.0g/L (TDS factor adjustable). Resistivity: 5.000Ω·cm~100.0MΩ·cm. Salinity: 0~100ppt |
| Conductivity: Calibration points | one point |
| Conductivity: Resolution | 0.001/0.01/0.1/1 |
| Conductivity: ATC range | (0.0~50.0)°C/(32.0~122.0)°F |
| Conductivity: Accuracy | ±0.5%F.S ±one digit |
| DO: Measurement range | (0.00~20.00)mg/L |
| DO: Accuracy | ±0.30mg/L |
| DO: DO saturation | (0.0~200.0)% |
| DO: Automatic barometric pressure compensation | (0~200.0)kPa |
| DO: Automatic temperature compensation | (0.0~50.0)°C/(32.0~122.0)°F |
| DO: Manual salinity compensation | 0~45ppt |
| Temperature: Measurement range | (0.0~100.0)°C/(32.0~212.0)°F |
| Temperature: Accuracy | ±0.2°C |
| Others: Charging socket | Type-C |
| Others: Power supply | Rechargeable lithium battery |
| Others: IP Rating | IP57 |
Keys, display icons and interfaces
| Element | Function |
|---|---|
| Charging socket | Type-C |
| Power key | Long press to power on/off |
| Hold icon (data locking) | Press to start data lock/free |
| Save icon (manual data logger) | (No., parameter, value, temp, date and time) |
| Timer icon (auto. timing data logger) | (0.5, 1, 2, 3, 5, 10, 20m) |
| Set icon (parameter setting key) | Parameter Setting Key |
| View icon | View data stored (press the up and down keys to display) |
| Mode icon (parameter selection) | pH > mV > ORP; COND (RES/TDS/SAL) > DO (SAT) |
| Down key | Parameter selection |
| CAL key | Press to enter calibration mode |
| ESC key | Press to return to measurement mode |
| Start/Enter key | Press to confirm calibration and parameter settings |
| SET key | Shortcut key to start parameter setting. In the pH measurement mode, long press to change the pH resolution: 0.1/0.01/0.001 |
| BNC socket | For pH/ORP probe |
| 4-pin socket | For temperature/cond probe |
| 8-pin socket | For DO probe |
| Silicone protective caps | Protect the sockets |
Package contents
| Supplied with the instrument | Qty |
|---|---|
| ALab™ DPCD10 meter | Included |
| Plastic ATC pH probe 201T-H | Included |
| Plastic ATC Cond probe CON201T, K=1.0 | Included |
| Polarographic DO probe DO01-2M-H, cable length: 2m | Included |
| pH buffer solution: pH4.00/7.00/10.01, 50mL, one for each | Included |
| Standard Cond solution: 84μS/1413μS/12.88mS/cm, 50mL, one for each | Included |
| Membrane refill solution: DOS-01, 30mL, 1btl | Included |
| Membrane sensor caps DOSC: 3pcs/group, 1pkg | Included |
| Rechargeable lithium battery (pre-installed) | Included |
| Charging cable | Included |
| Hard carrying case (L/W/H: 36.0×32.0×9.0cm) | Included |
Package contents reflect the current published configuration; the itemized bill of materials is confirmed on the quotation.
Where it fits
Frequently asked questions
Which parameters does the DPCD10 measure?
The DPCD10 measures pH/mV, mV (ORP), conductivity (with TDS, salinity and resistivity), dissolved oxygen and temperature.
Which probes are included?
A plastic ATC pH probe 201T-H, a plastic ATC Cond probe CON201T (K=1.0) and a polarographic DO probe DO01-2M-H (cable length: 2 m); see Package contents for the solutions and the carrying case.
How many readings can be stored?
The meter stores up to 3000 groups of data. It has a manual data logger (No., parameter, value, temp, date and time) and an auto. timing data logger (0.5, 1, 2, 3, 5, 10, 20m).
Which sockets does the meter have?
A BNC socket for the pH/ORP probe, a 4-pin socket for the temperature/cond probe and an 8-pin socket for the DO probe, with silicone protective caps.
How is the meter powered and what is its protection class?
It has a rechargeable lithium battery, a Type-C charging socket and protection class IP57.
How can the pH resolution be changed?
In the pH measurement mode, long press the SET key to change the pH resolution: 0.1/0.01/0.001.