ALab™ C410 Chemistry-Design Diaphragm Vacuum Pump
Product Detail
ALab™ C410 Chemistry-Design Diaphragm Vacuum Pump is an oil-free laboratory vacuum pump. The manufacturer specifies it as the vacuum source for the RE100-Pro and RE100-S distillation systems; it is also used for general oil-free laboratory vacuum duties such as filtration, degassing and vacuum drying, where application suitability depends on vapor compatibility, system configuration and the use of appropriate traps. It reaches an ultimate vacuum of 13 mbar, moves 25 L/min of free air, and is rated for continuous operation with a manufacturer-stated noise level of 50 dB.

Why choose the C410
Match the pump to your system
A vacuum pump has no single “speed”: capacity falls as the pressure drops. Set your system volume and target pressure to read the interpolated pumping speed and an idealized pump-down time from atmosphere, within the 280–950 mbar range covered by the manufacturer’s published data:
Model: the curve is a linear interpolation of the six manufacturer-published pressure/flow pairs (1000–280 mbar). Below 280 mbar the manufacturer publishes no pumping-speed data, so no values are estimated there — the 13 mbar figure is the ultimate-vacuum rating, not a speed. Times integrate the throughput balance t = V ∫ dp / [S(p) · p] from 1000 mbar for a closed, leak-tight, isothermal system; real systems are slower once hose conductance, traps, leaks and evaporating solvent vapor are added. A preliminary teaching tool, not sizing software.
Product specifications
| Parameter | C410 |
|---|---|
| Brand | ALab™ by ACS Material |
| SKU # | ALC41000 · variants ALC410US / ALC410EU / ALC410UK |
| Ultimate Vacuum (Final Pressure) | 13 mbar |
| Free Air Capacity | 25 L/min |
| Work Mode | Continuous operation |
| Motor Speed | 1450 rpm |
| Nozzle Size | 8 mm |
| Power | 95 W |
| Noise | 50 dB |
| Flow Rate — 1000 mbar | 25 L/min |
| Flow Rate — 880 mbar | 22 L/min |
| Flow Rate — 750 mbar | 20 L/min |
| Flow Rate — 600 mbar | 12 L/min |
| Flow Rate — 480 mbar | 7.5 L/min |
| Flow Rate — 280 mbar | 5 L/min |
| Voltage / Frequency | 110 V / 60 Hz (US) · 220 V / 50 Hz (EU, UK) |
Ordering options · choose voltage, frequency and plug configuration
The C410 is supplied in one electrical configuration per unit. Confirm the mains supply of the destination laboratory before ordering. Performance figures on this page are the manufacturer’s published values for the model; version-specific performance for the 110 V / 60 Hz configuration is pending manufacturer confirmation.
| Configuration | SKU # | Voltage / Frequency |
|---|---|---|
| ALab™ C410 · US plug | ALC410US | 110 V / 60 Hz |
| ALab™ C410 · EU plug | ALC410EU | 220 V / 50 Hz |
| ALab™ C410 · UK plug | ALC410UK | 220 V / 50 Hz |
Pumping speed at a glance
| Pressure (mbar) | 1000 | 880 | 750 | 600 | 480 | 280 |
|---|---|---|---|---|---|---|
| Pumping speed (L/min) | 25 | 22 | 20 | 12 | 7.5 | 5 |
Only the listed pressure/flow pairs are manufacturer-published; intermediate pumping speeds are not specified. Below 280 mbar no speed data is published — 13 mbar is the ultimate-vacuum rating of the pump.
Built for the RE100 rotary evaporators. The C410 is the matched vacuum source for the RE100-Pro and RE100-S. A programmable vacuum controller can be added between pump and condenser for controlled pressure ramps that can help reduce bumping, together with the vacuum regulator / moisture trap (Cat. No. 18900186) for fine manual trimming and moisture protection. Use an appropriate cold trap or moisture trap when handling condensable vapors, and route solvent-laden exhaust in accordance with your laboratory’s safety procedures.
FAQ
Is 13 mbar deep enough for rotary evaporation?
For routine work with common solvents, yes: ethanol boils near 40 °C at about 178 mbar and water near 40 °C at about 74 mbar, both well above 13 mbar. Note that 13 mbar is the pump’s ultimate-vacuum rating, not a pressure it will hold during evaporation: actual operating pressure depends on vapor load, condenser performance, tubing conductance, system tightness and solvent compatibility. Very high-boiling solvents and short-path work call for a deeper-vacuum pump.
Why a diaphragm pump instead of a rotary-vane pump?
Solvent vapor passes through the pump head during evaporation. In an oil-sealed rotary-vane pump that vapor can dissolve or condense in the oil, reducing oil life and achievable vacuum and potentially carrying solvent-contaminated oil mist into the exhaust; a chemistry-design diaphragm pump has no oil to contaminate and needs no oil changes.
How do I know the pump is big enough for my system?
Compare the speed at your working pressure with the volume you have to evacuate. The tool on this page provides a preliminary idealized comparison based on the manufacturer-published pumping-speed data; allow generous margin over it, because hoses, traps and the solvent vapor load all reduce effective speed.
Can it run all day?
The pump is rated for continuous operation and has a manufacturer-stated noise level of 50 dB with a 95 W motor. Provide free airflow around the housing and keep the exhaust unobstructed.
Which electrical configuration should I order?
US is 110 V / 60 Hz; EU and UK are 220 V / 50 Hz. One voltage, frequency and plug configuration is supplied per unit — check the mains supply at the destination laboratory before ordering.
What hose does it need?
The nozzles are 8 mm. Use vacuum tubing rated for the solvents in use, keep runs short and bores wide, and place a cold trap or moisture trap ahead of the pump when condensable vapor load is high.
Related
ALab™ is a trademark of ACS Material, LLC. Disclaimer: ACS Material, LLC believes the information on this page is accurate and represents the best and most current information available to us. Specifications reflect the manufacturer's current documentation; final configurations and accessories are confirmed at quotation — always refer to the model datasheet for guaranteed ratings, and follow your laboratory's safety procedures for vacuum operation and solvent handling. The interactive simulator is a schematic teaching tool, not sizing software or performance data for a specific unit. ACS Material makes no representations or warranties, express or implied, regarding suitability for any purpose, and will not be responsible for damages resulting from use of or reliance upon this information.