GEt Quote

ALab™ DHP4-550 LCD Digital 4-Inch Hotplate

In stock
SKU# ALDH010F

Product Detail

The ALab™ DHP4-550 is an LCD digital hotplate with a 4-inch (100 × 100 mm) glass-ceramic plate that heats from RT to 550°C. It has no stirring function; for heating and stirring on the same plate size, see the DMS4-550.

ALab™ DHP4-550 LCD Digital 4-Inch Hotplate with a square glass-ceramic plate and LCD, shown with a support stand and temperature probe
ALab™ DHP4-550 LCD Digital 4-Inch Hotplate — image shown for reference (same photo as the DMS4-550). The DHP4-550 has no stirring function.

Three heating modes (A rapid, B standard, C stable) suit different samples. The LCD shows the set and actual temperature and the time, the timer runs from 1 min to 99 h 59 min and stops heating at the end of the countdown, and an external PT1000 temperature sensor (accuracy ±0.2°C) controls the liquid temperature.

580°C overheating protection, two independent safety circuits, a settable safety temperature and a "HOT" warning that flashes while the plate is above 50°C, even after the power is switched off, protect the user. A safety lock, a buzzer and an RS232 connector for remote control from a PC are built in.

Why choose the DHP4-550

4-inch glass-ceramic plateResistant to strong acids and alkalis and high temperatures; easy to clean.
RT to 550°CPlate heating in 1°C increments; control accuracy ±1°C below 100°C and ±1% above.
Three heating modesA rapid, B standard and C stable heating.
PT1000 controlExternal temperature control with a PT1000 sensor, accuracy ±0.2°C.
Safety580°C overheating protection, two safety circuits, "HOT" warning above 50°C.
Timer and RS2321 min to 99 h 59 min; remote control from a PC.
Plate
100 × 100 mm · 4 in
Temperature
RT to 550°C
Heating output
410 W
Timer
to 99 h 59 min
Data
RS232

Specifications General — DHP4-550

ParameterSpecification
BrandALab™ by ACS Material
ModelDHP4-550
Product typeLCD digital hotplate with 4-inch glass-ceramic plate (no stirring)
Work plate dimension100 × 100 mm (4 inch)
Work plate materialGlass ceramic
Temperature displayLCD
Heating temperature rangeRT to 550°C, increment 1°C
Temperature control accuracy (liquid)±1°C (<100°C), ±1% (>100°C)
Overheating protection580°C
Temperature display accuracy0.1°C
External temperature sensorPT1000 (accuracy ±0.2°C)
"HOT" warning50°C
Data connectorRS232
Protection classIP21
Timer function1 min to 99 h 59 min
Power430 W
Heating output410 W
Voltage, frequency100 to 120 V / 200 to 240 V, 50/60 Hz
Dimensions [L × W × H]270 × 175 × 108 mm
Weight1.8 kg
Permissible ambient temperature and humidity5 to 40°C, 80% RH

Control and safety

FunctionDetail
Heating modesA: rapid heating, B: standard heating, C: stable heating
Timer1 min to 99 h 59 min; heating stops at the end of the countdown
Overheating protectionOverheating protection at 580°C
Two independent safety circuitsContinuous monitoring: on overheating or sensor failure the unit reports an error and stops heating; on a short circuit it cuts the power supply
"HOT" warningWhen the plate is above 50°C and heating is off, "HOT" flashes on the display, even after the power is switched off
Safety temperatureSettable safety temperature limit
External temperature controlPT1000 temperature sensor, accuracy ±0.2°C, for control of the liquid temperature
Safety lockLocks the settings against accidental changes during operation
BuzzerSounds when the unit detects a fault and when the timer countdown ends
RS232 data connectorRemote control from a PC

Order codes ACS Cat. No.

Model / configurationACS Cat. No.
DHP4-550 LCD Digital 4-Inch HotplateALDH0100

Parent ACS Cat. No.: ALDH010F. Order ACS Cat. No.: ALDH0100.

Where it fits

The DHP4-550 heats samples from RT to 550°C on a compact 4-inch glass-ceramic plate, without stirring. For heating and stirring, see the DMS4-550; for a larger 7-inch hotplate, see the DHP7-550.

Frequently asked questions

Does the DHP4-550 stir?

No. The DHP4-550 is a hotplate without a stirring motor. The DMS4-550 combines heating to 550°C with magnetic stirring on the same plate size.

What is the heating range?

From RT to 550°C in 1°C increments, with 580°C overheating protection. The heating output is 410 W.

Can I control the liquid temperature?

Yes. External temperature control is possible with a PT1000 sensor (accuracy ±0.2°C); the control accuracy is ±1°C below 100°C and ±1% above 100°C.

Does it have a timer?

Yes. The timer runs from 1 min to 99 h 59 min; heating stops at the end of the countdown.

What happens when the plate is still hot?

When the plate is above 50°C and heating is off, "HOT" flashes on the display, even after the power is switched off.

ALab™ DMS4-550 Magnetic Hotplate Stirrer — 4-inch LCD magnetic hotplate stirrer, RT to 550°C, 10 L.
ALab™ DHP7-550 LCD Digital 7-Inch Hotplate — 7-inch glass-ceramic hotplate, RT to 550°C, timer, RS232; no stirring.
ALab™ DMS4 LCD Digital 4-Inch Magnetic Stirrer — 100 × 100 mm glass-ceramic plate, 50 to 1500 rpm, 10 L, timer, RS232; no heating.
ALab™ DMS5-340 LCD Digital 5-Inch Timed Heating Magnetic Stirrer — Ø135 mm plate, RT to 340°C, 50 to 1500 rpm, 20 L, uniform magnetic field.
ALab™ is a trademark of ACS Material, LLC. Disclaimer: ACS Material, LLC believes the information on this page is accurate and represents the best and most current information available to us. Specifications reflect the current published data; refer to the model-specific specifications and follow your laboratory's safety procedures for hotplate operation. ACS Material makes no representations or warranties, express or implied, regarding suitability for any purpose, and will not be responsible for damages resulting from use of or reliance upon this information.