ALab™ dFlow-Pro Electronic Bottle-Top Dispenser
Product Detail
The ALab™ dFlow-Pro is a touchscreen electronic bottle-top dispenser with multiple dispensing modes, in three sizes: 10 mL, 25 mL and 50 mL. Dispensing accuracy complies with ISO 8655-5.

A 4.3-inch TFT color touchscreen sets the dispensing programs, aspiration and dispense speed is adjustable in 10 levels for different viscosities, and six modes cover manual and automatic fixed-volume dispensing, multi-step and custom programs, free dispensing and titration.
The consumable-free, low-maintenance design has a built-in 0 to 90° rotary valve with quick venting and reagent recirculation, a standard drying tube for moisture- and CO2-sensitive reagents, and transparent and amber windows as standard. The dFlow-Pro connects to a PC over Type-C and WiFi and stores 1000 experiment records.
Why choose the dFlow-Pro
Specifications General — dFlow-Pro 10 mL, 25 mL, 50 mL
| Parameter | Specification |
|---|---|
| Brand | ALab™ by ACS Material |
| Model | dFlow-Pro |
| Product type | Electronic bottle-top dispenser |
| Sizes | 10 mL, 25 mL, 50 mL |
| Aspiration / dispense speed | Adjustable, 10 levels |
| Dispensing modes | Manual fixed-volume, automatic fixed-volume, automatic multi-step (up to 10 steps), manual free, custom, titration |
| Rotary valve | Built-in, 0 to 90°, quick venting and reagent recirculation |
| Windows | Transparent and amber (standard) |
| Drying tube | Standard |
| Operating temperature range | 10 to 30°C |
| Quality standard | ISO 8655 (dispensing accuracy per ISO 8655-5) |
| Display | 4.3-inch TFT color touchscreen, 800 × 480 resolution |
| Memory | 1000 experiment records |
| PC interface | Type-C and WiFi |
| Bluetooth function | Yes |
| WiFi function | Yes |
| Battery capacity | 3500 mAh |
| Weight | 1.2 kg |
Sizes 10 mL, 25 mL, 50 mL
| Size | ACS Cat. No. |
|---|---|
| 10 mL | ALDFLP01 |
| 25 mL | ALDFLP02 |
| 50 mL | ALDFLP03 |
Working accuracy by measured volume
| Size | Measured volume (mL) | Systematic error (µL) | Systematic error (%) | Random error (µL) | Random error (%) |
|---|---|---|---|---|---|
| 10 mL | 10 | ±50 | ±0.5 | ±30 | ±0.3 |
| 5 | ±50 | ±1.0 | ±30 | ±0.6 | |
| 1 | ±50 | ±5.0 | ±30 | ±3.0 | |
| 25 mL | 25 | ±125 | ±0.5 | ±62.5 | ±0.25 |
| 12.5 | ±125 | ±1.0 | ±62.5 | ±0.5 | |
| 2.5 | ±125 | ±5.0 | ±62.5 | ±2.5 | |
| 50 mL | 50 | ±250 | ±0.5 | ±125 | ±0.25 |
| 25 | ±250 | ±1.0 | ±125 | ±0.5 | |
| 5 | ±250 | ±5.0 | ±125 | ±2.5 |
Dispensing accuracy complies with ISO 8655-5.
Dispensing modes six modes
| Dispensing mode | How it works |
|---|---|
| Manual fixed-volume dispensing | Set the volume (mL); aspiration and dispensing are controlled manually. |
| Automatic fixed-volume dispensing | Set the dispensing interval, volume and repetitions; trigger the first dispense manually, then the program continues at the set interval. |
| Automatic multi-step dispensing | Set volumes for up to 10 steps (total not exceeding the piston capacity) plus the interval time; trigger the first dispense manually, then the program runs at the set intervals. |
| Manual free dispensing | Aspiration and dispensing volumes are controlled manually. |
| Custom dispensing | Define steps, intervals, volumes, aspiration and dispensing, and save them as a program for future use. |
| Titration | Manual or continuous titration until stopped, with the titration volume displayed (e.g., volumetric flask preparation, pH adjustment). |
Order codes one ACS Cat. No. per configuration
| Model / configuration | ACS Cat. No. |
|---|---|
| dFlow-Pro 10 mL | ALDFLP01 |
| dFlow-Pro 25 mL | ALDFLP02 |
| dFlow-Pro 50 mL | ALDFLP03 |
Parent ACS Cat. No.: ALDFLP. Select the configuration in the order menu.
Package contents
| Supplied with the instrument | Qty |
|---|---|
| dFlow-Pro electronic bottle-top dispenser of the selected size | 1 |
| Drying tube | Included |
| Transparent and amber windows | Included |
Where it fits
Frequently asked questions
Which sizes are available?
Three sizes: 10 mL, 25 mL and 50 mL.
How accurate is the dFlow-Pro?
At the nominal volume the systematic error is ±0.5% and the random error ±0.3% (10 mL) or ±0.25% (25 and 50 mL); dispensing accuracy complies with ISO 8655-5. The values per measured volume are in the accuracy table.
Which dispensing modes does it have?
Manual fixed-volume, automatic fixed-volume, automatic multi-step (up to 10 steps), manual free, custom dispensing with saved programs, and titration.
Can it handle light-sensitive reagents?
Yes. Transparent and amber windows are standard; the amber window suits light-sensitive liquids such as iodine solution, potassium permanganate and silver nitrate.
Does it need consumables?
The dispenser has a consumable-free design. For moisture- or CO₂-sensitive reagents, users can add a compatible absorbent to the standard drying tube.
How does it connect to a PC?
Over Type-C and WiFi; Bluetooth is also available, and the memory stores 1000 experiment records.