ALab™ RE200-Pro 20 L Industrial Digital Rotary Evaporator
Product Detail
ALab™ RE200-Pro 20 L Industrial Digital Rotary Evaporator is a pilot-scale rotary evaporation system for large-volume distillation, separation, sample concentration and solvent recovery. A large LCD shows heating temperature, rotation speed, direction and time; one-click motorized lifting raises the 20 L evaporating flask over a 180 mm stroke; and a two-section vertical triple-coil condenser with 1.2 m² of cooling surface supports up to 4.5 L/h on 75% alcohol.

Why choose the RE200-Pro
Find your working vacuum
A rotary evaporation system lives on the interplay of pressure and temperature: lower the system pressure and the solvent boils at a gentler temperature. Pick your solvent and drag the pressure to see where the boiling point lands:
Model: Antoine vapor-pressure equation lg P = A − B/(C + T) for pure solvents within published coefficient ranges; the bath and condenser guides follow the Δ20 °C (20/40/60) rule of thumb. The 13 mbar reference line in the tool is the C410 pump’s ultimate-vacuum rating. The RE200-Pro has a manufacturer-stated ultimate-vacuum rating below 2.6 mbar; a deeper-vacuum pump is required only when the intended process must approach that limit, and final pump selection must also account for pumping speed and vapor load. The tool computes ideal boiling points for pure solvents — it does not predict the evaporation rate of a 20 L system and is not a pump-selection tool. Curves are schematic teaching aids, not performance data for a specific solvent lot or system.
Product specifications
| Parameter | RE200-Pro |
|---|---|
| Brand | ALab™ by ACS Material |
| SKU # | ALRE2P20 |
| Temperature Range | Room temperature ~ 180 °C (water and oil bath) |
| Temperature Control | Microprocessor PID control · Water: ±1 °C · Oil: ±3 °C |
| Rotation Speed | 10–150 rpm · clockwise / counterclockwise · interval operation |
| Ultimate Vacuum (Final Pressure) | Less than 2.6 mbar |
| Evaporation Capacity | Max. 4.5 L/h (75% alcohol) |
| Display | LCD (temperature / speed / direction / time) |
| Time Setting Range | 0–999 min |
| Lift Stroke | Motorized, 180 mm · one-click operation |
| Safety Features | Motor over-current · earth-leakage · lifting-overload · boil-dry · overheating protection |
| Evaporating Flask | Round flask, 20 L |
| Receiving Flask | Round flask, 10 L, with drain valve |
| Condenser | Two-section vertical triple-coil · cooling surface 1.2 m² |
| Heating Bath Size | Ø 450 × 240 mm |
| Hose Connection | Cooling / suction nozzle OD 16 mm · vacuum pump nozzle OD 16 mm |
| Dimensions [D×W×H] | 1160 × 600 × 1860 mm |
| Voltage / Frequency | 220 V, 50/60 Hz |
| Total Power / Heating Power | 4600 W / 4400 W |
Package contents
| Supplied with the system | Qty |
|---|---|
| Main unit with motorized lift, 180 mm | 1 |
| Water / oil heating bath, Ø 450 × 240 mm | 1 |
| Vertical glassware set 18203071 — configuration as confirmed on the quotation, including the specified two-section triple-coil condenser (1.2 m²) and flask assembly (20 L evaporating, 10 L receiving with drain valve) | 1 |
The itemized bill of materials of the glassware set, together with clamps, seals and connectors, is confirmed on the quotation; the manufacturer does not publish the set contents in the catalogue. Vacuum pump, recirculating chiller and vacuum-control options are ordered separately and matched to the process.
Pump, recirculating chiller and vacuum-control options are system-level configuration items. They are selected separately and confirmed at quotation according to the intended process. The manufacturer states an ultimate vacuum below 2.6 mbar for the RE200-Pro; the benchtop C410 diaphragm pump is rated to 13 mbar and is specified by the manufacturer for the RE100-Pro and RE100-S, not for this system. Selection is made against your solvents, batch size and target working pressure — together with the electrical connection (single or three phase, rated current, breaker size, plug or hard-wiring and supply cord).
FAQ
What vacuum pump does the RE200-Pro need?
The manufacturer states an ultimate vacuum of less than 2.6 mbar for the RE200-Pro. Vacuum-pump selection depends on the solvent, vapor load, desired working pressure, pumping speed and complete system configuration; if the process must approach the published ultimate-vacuum limit, a suitably deeper-vacuum pump is required. Tell us your solvents, batch size and target working pressure and the pump is matched at quotation.
How does it compare with the benchtop RE100 models?
The RE100-Pro and RE100-S take flasks up to 3 L and evaporate up to 1.56 L/h of water. The RE200-Pro takes a 20 L flask and reaches 4.5 L/h on 75% alcohol, with a 1.2 m² condenser instead of 1200–1700 cm². Method development stays on the bench; the RE200-Pro carries the process into 20 L pilot batches, with parameters re-optimized for the larger vapor load and condenser duty.
Can distillate be collected without stopping the run?
Yes — the 10 L receiving flask has a drain valve, and the switch-valve arrangement supports continuous distillate removal without breaking the system vacuum.
What space and services does it need?
The unit measures 1160 × 600 × 1860 mm on a mobile frame and runs on 220 V, 50/60 Hz with a total power of 4600 W. Allow clearance above the condenser column for glassware handling, and plan cooling-water or chiller connections at the 16 mm hose nozzles.
Water bath or oil bath?
The bath supports water or oil heating over a manufacturer-stated overall range from room temperature to 180 °C, under microprocessor PID control with ±1 °C accuracy in water and ±3 °C in oil. Water is used for lower-temperature operation, while a suitable heat-transfer oil is required for higher temperatures — confirm medium-specific operating limits in the operating manual. The system includes boil-dry and overheating protection.
How do I choose bath temperature and vacuum together?
Start from the Δ20 °C rule: set the vacuum so the solvent boils about 20 °C below the bath, and keep the condenser about 20 °C below the vapor temperature. The interactive tool on this page computes it per solvent; at 20 L scale, confirm the result against your own process and condenser capacity.
Related
ALab™ is a trademark of ACS Material, LLC. Disclaimer: ACS Material, LLC believes the information on this page is accurate and represents the best and most current information available to us. Specifications reflect the manufacturer's current documentation; final configurations and accessories are confirmed at quotation — always refer to the model datasheet for guaranteed ratings, and follow your laboratory's safety procedures for vacuum operation, heated baths and solvent handling. The interactive simulator is a schematic teaching tool, not performance data for a specific unit. ACS Material makes no representations or warranties, express or implied, regarding suitability for any purpose, and will not be responsible for damages resulting from use of or reliance upon this information.