A useful lab equipment list starts with how a sample moves across the bench, not with a catalog. In most research labs a sample is measured out, mixed or heated, separated, measured again, and kept at a set temperature along the way. This guide follows that path step by step. It explains what each class of benchtop instrument does and which specifications matter when you choose one. It ends with editorial checklists for four common lab types, each item marked Essential, Common or Can wait.
Quick answer: For most new research labs, the core benchtop equipment list has five groups. (1) Liquid handling: pipettes, a pipette controller and bottle-top dispensers. (2) Mixing and heating: magnetic hotplate stirrers, a vortex mixer and, where needed, an overhead stirrer or homogenizer. (3) Separation: a microcentrifuge and, for solvent work, a rotary evaporator with a vacuum source. (4) Measurement: a pH/conductivity meter and a UV-Vis spectrophotometer. (5) Temperature control: dry baths, a chiller or cooling block. Molecular biology labs add a thermal cycler, electrophoresis and gel imaging. Buy the instrument that limits your daily workflow first; the rest can follow as throughput grows.

1. Plan Before You Buy: Map the Workflow
Equipment chosen one item at a time often leaves gaps: a hotplate with no way to monitor the solution temperature, or a centrifuge whose rotor does not take the tubes the lab actually uses. A short planning exercise before you write a lab startup equipment request avoids most of these mismatches.
Map the sample path
Write down your three to five most frequent procedures as a sequence of steps. For each step, note the volume handled, the temperature range, the solvents or reagents involved and how many samples move through it per day. A typical materials chemistry synthesis, for example, might read: dispense precursor solution, stir and heat to reflux, cool, centrifuge to collect particles, redisperse, measure absorbance. Each verb maps to a class of equipment, and the numbers attached to it (volume, temperature, sample count) map to the specifications you need.
| Step in the sample path | Equipment class | Specifications that usually decide the choice |
|---|---|---|
| Measure and transfer liquids | Pipettes, dispensers, burettes | Volume range, number of channels, chemical compatibility, autoclavability |
| Mix, heat and disperse | Hotplate stirrers, overhead stirrers, homogenizers, vortex mixers, shakers | Plate size and material, maximum temperature, speed range, stirring volume, viscosity limit |
| Separate | Centrifuges, rotary evaporators, vacuum pumps | Maximum RCF (×g), rotor and tube formats, refrigeration, flask size, vacuum level |
| Measure | UV-Vis spectrophotometers, pH/conductivity/DO meters | Wavelength range, bandwidth, measurement parameters, calibration and temperature compensation |
| Hold temperature | Dry baths, thermo mixers, cooling blocks, chillers | Temperature range, stability, block formats, cooling capacity |
Size by throughput, not by aspiration
A lab that runs ten samples a week has different needs from one that runs a hundred a day. Throughput affects three choices more than any others:
- Channels and formats. If plates or tube strips are routine, a multichannel pipette and a microcentrifuge rotor or spin-down unit that accepts strips will save more time than a faster single-channel instrument.
- Positions. Parallel reactions favor a multi-position stirrer over several single hotplates sharing a bench.
- Duty cycle. Instruments that run all day, such as a chiller feeding a condenser, should be specified for continuous use at the temperatures you actually need.
Buy the bottleneck first
When you cannot equip everything at once, rank each item by how often a missing or shared instrument stops work. A shared pipette set or a single hotplate becomes a queue within a week; a second spectrophotometer rarely does. Items that are used by every person every day belong at the top of the list. Items used for one technique a month can usually be shared with a neighboring lab or added later. The checklists in Section 8 follow this logic.
Write the safety plan alongside the list
Equipment decisions and safety decisions are linked. In the United States, the OSHA Laboratory Standard requires that "where hazardous chemicals as defined by this standard are used in the workplace, the employer shall develop and carry out the provisions of a written Chemical Hygiene Plan" [1]. That plan must include "a requirement that fume hoods and other protective equipment are functioning properly and specific measures that shall be taken to ensure proper and adequate performance of such equipment" [1]. Planning ventilation, electrical circuits and waste handling together with the equipment list is easier than retrofitting them later. Prudent Practices in the Laboratory from the National Research Council is a widely used reference for this planning [2]. Biological work should also be planned around a protocol-specific risk assessment, which the CDC/NIH BMBL describes as its "core principle" [3].
2. Liquid Handling: Pipettes, Dispensers and Burettes
Almost every procedure starts with moving a measured volume of liquid, so liquid handling is usually the first purchase and the one most often under-specified. Volume errors carry through every downstream step.
Pipettes
Air-displacement piston pipettes cover the microliter range and are the default tool for aqueous solutions. Requirements and test methods for piston pipettes are set out in ISO 8655-2, and the gravimetric reference procedure used to check them is described in ISO 8655-6 [4][7]. When building a pipette set, consider:
- Volume coverage. A set of three or four single-channel pipettes typically covers 0.5 µL to 1000 µL, with a larger-volume pipette added for 1–10 mL transfers. Pipette performance varies by model and set volume. Check the manufacturer's accuracy and precision specifications at the volumes you expect to dispense; overlapping ranges can help you choose a suitable pipette for each task.
- Mechanical or electronic. Mechanical pipettes are simple and robust. Electronic pipettes add motor-driven aspiration and programmed modes (such as repeat dispensing or mixing), which reduce variation between users and hand strain during long runs.
- Autoclavability. Molecular and cell biology labs often need pipettes whose lower section, or the whole body, can be autoclaved.
ALab™ options include the Toppette mechanical pipette, offered in single-channel, 8-channel, 12-channel and fixed-volume versions; its single-channel models cover 0.1 µL to 10000 µL across 13 ranges, and its lower section is autoclavable. The MicroPette Plus is fully autoclavable at 121 °C. For electronic pipetting, the dPette and dPette+ cover 0.5 µL to 1000 µL in single-channel models, and the dPette+ single-channel adds pipetting, mixing, stepper and dilution modes. The full range is on the pipettes category page.
Pipette controllers
Serological pipettes in the 1–100 mL range need a pipette controller. Look for speed control, a replaceable hydrophobic filter that protects the instrument from liquid ingress, and a rechargeable battery. The Levo Plus pipette controller works with serological pipettes from 0.1 mL to 100 mL, has six speeds and uses a 0.45 µm hydrophobic filter.
Bottle-top dispensers and burettes
Repeated dispensing of the same reagent, such as a solvent, an acid or a buffer, is faster and cleaner from a bottle-top dispenser than from a graduated cylinder. ISO 8655-5 covers dispensers, and ISO 8655-3 covers piston burettes [5][6]. The main selection factor is chemical compatibility of the wetted parts with your reagents.
| Need | What to look for | ALab™ example |
|---|---|---|
| General reagents, 0.5–100 mL | Several volume ranges, autoclavable, reagent recovery | DispensMate-Pro: 0.5–100 mL in five ranges; fully autoclavable at 121 °C |
| Hydrofluoric acid | Wetted parts resistant to HF (glass is attacked) | DispensMate-HF: 1.0–10 mL; ceramic piston, cylinder and valve; FEP tubes |
| Electronic, programmable dispensing | Motor drive, speed settings, stepper mode | dFlow: 0.1–99.9 mL; 16 speeds; continuous or stepper mode |
| Titration | Fine resolution, integrated stirring | dTrite electronic burette: 0.01–99.99 mL; 10 µL resolution; built-in magnetic stirrer |
| Repetitive small aliquots | Positive displacement, many steps per fill | StepMate: 10 µL–5 mL per step; up to 48 steps per filling |
HF requires dedicated training, PPE and emergency procedures well beyond dispenser choice; consult the SDS and the NIOSH Pocket Guide entry before any HF work [20]. See all models on the bottle-top dispensers page.
Liquid waste aspiration
Cell culture and molecular biology workflows may generate liquid waste that can be collected with a vacuum aspiration system configured for the waste and its hazards. Appropriate collection traps, filters, decontamination and handling procedures should follow the laboratory's risk assessment and applicable biosafety requirements. The SafeVac vacuum-controlled aspiration system has a 0–600 mbar vacuum range, a 4 L collection bottle and level detection to prevent overfilling. Related units are on the vacuum aspirators page.
3. Stirring, Heating and Dispersing
The second stage of most sample paths is mixing, often with heat. The right instrument depends on volume, viscosity and the temperature the chemistry needs.
Magnetic hotplate stirrers
A magnetic hotplate stirrer is a common choice for chemistry benches where solutions need to be mixed and heated. When comparing models, focus on:
- Plate material. Glass-ceramic plates tolerate high temperatures and many chemicals. Coated metal plates spread heat quickly and evenly over a smaller area.
- Maximum temperature and control. Check whether the instrument accepts an external temperature sensor placed in the liquid. Controlling to the solution temperature, rather than the plate temperature, gives more repeatable results.
- Safety features. Independent overheat protection and a residual-heat warning are worth checking on every model.
- Stirring volume. Manufacturers usually quote this for water; viscous mixtures stir less effectively.
| Model | Plate | Temperature range | Speed | Notes from the product page |
|---|---|---|---|---|
| MS-H-Pro+ | 5-inch (Ø135 mm) ceramic-coated stainless steel | RT to 340 °C | 100–1500 rpm | 20 L (H2O) stirring volume; PT1000 external sensor; overheat protection at 420 °C; "HOT" warning above 50 °C |
| MS7-H550-Pro | 7-inch (184 × 184 mm) glass-ceramic | RT to 550 °C | 100–1500 rpm | 20 L stirring volume; PT1000 sensor; overheat protection at 580 °C |
| MS-H-S10 | 10 positions on a 180 × 450 mm plate | RT to 120 °C | 0–1100 rpm | 0.4 L per position; parallel reactions and screening |
| MS-H280-B / F series | Heating mantle with stirring | RT to 280 °C | 200–1500 rpm | 100–1000 mL flat-bottom beakers (B) or round-bottom flasks (F); PT1000 sensor |
For heating without stirring, the DHP7-550 hotplate has a 7-inch glass-ceramic plate, reaches 550 °C and offers three heating modes. For stirring at room temperature, the non-heating MS-S magnetic stirrer stirs up to 20 L. The magnetic stirrers and hotplates category pages list the full ranges.
Overhead stirrers
Magnetic coupling weakens as viscosity rises. Polymer solutions, slurries and pastes, or volumes beyond what a stir bar can move, need an overhead stirrer that drives a shaft and impeller directly. The key specifications are maximum viscosity, maximum torque and stirring volume. In the OS-Pro series, the OS20-Pro, OS40-Pro and OS70-Pro are rated for 20, 40 and 70 L, with maximum viscosities of 10000, 50000 and 100000 mPa·s. Each displays speed and torque, which helps track viscosity changes during a reaction. More options are on the overhead stirrers page.
Homogenizers and dispersers
Breaking up agglomerates, making emulsions or dispersing nanoparticles in a liquid calls for high shear. A rotor–stator homogenizer forces the sample through a narrow gap at high speed. The D-500 high-shear homogenizer runs at 10000–30000 rpm and handles working volumes from 10 mL to 40000 mL, depending on the dispersing head. For small samples, the D-160 handheld homogenizer covers 0.1–250 mL at 8000–32000 rpm. See the homogenizers page.
Vortex mixers, rollers and shakers
Tube-scale mixing is usually handled by a vortex mixer, a tube roller or a shaker:
- Vortex mixer. Resuspends pellets and mixes small tubes in seconds. The MX-S vortex mixer runs at 0–3000 rpm with touch or continuous operation.
- Tube roller. Gives gentle, continuous mixing for viscous reagents or particle suspensions that should not foam. The MX-T6-S tube roller has six rollers turning at 0–70 rpm.
- Orbital shaker. Agitates flasks, plates or staining trays for long periods. The SK-O series orbital shakers include five models, with timers from 1 minute to 99 h 59 min.
The mixers category page groups vortex mixers, rollers and shakers.
4. Separation: Centrifuges and Rotary Evaporators
After mixing and reaction, most samples need something removed: a solid from a liquid, or a solvent from a product.
Centrifuges
The most important centrifuge specification is relative centrifugal force (RCF, expressed as ×g), not rpm. RCF depends on both rotor speed and rotor radius, so two instruments with the same rpm can deliver different forces. Protocols should specify ×g; check that the instrument and rotor reach it. Next, match the rotor to your tubes: 0.2 mL PCR tubes and strips, 1.5/2 mL microtubes, 15/50 mL conical tubes or larger bottles. Refrigeration matters for temperature-sensitive biological samples. Centrifuge safety, including balancing loads and inspecting rotors, is covered in Prudent Practices [2].
| Use | ALab™ example | Published figures |
|---|---|---|
| Quick spin-down of tubes and strips | D1008 palm micro-centrifuge | 7000 rpm, 2680 ×g; 8 × 0.2–2 mL tubes or PCR strips |
| Routine microtube work | D3024 high-speed micro-centrifuge | 15000 rpm, 21380 ×g; 24 × 1.5/2 mL rotor; dual door interlock |
| Temperature-sensitive microtube work | D1524R refrigerated micro-centrifuge | 15000 rpm, 21380 ×g; −20 to 40 °C; pre-cools to 4 °C in 7 min |
| Larger volumes, particle collection | DG1616 / DG1616R high-speed centrifuges | 16000 rpm, 28621 ×g; up to 4 × 400 mL; R model −10 to 40 °C |
The centrifuge category page lists every model and rotor.
Rotary evaporators and vacuum
For labs that regularly remove solvent from heat-sensitive samples or recover solvent, a rotary evaporator may be useful. Whether it is appropriate depends on the solvent, sample sensitivity, batch size, throughput and available alternatives. A rotary evaporator lowers the boiling point under vacuum, spreads the liquid as a thin film on a rotating flask and condenses the vapor for recovery. Our complete guide to rotary evaporation covers principles, setup and solvent recovery in depth. Shorter explainers cover how a rotary evaporator works in five steps, why the flask rotates, solvent boiling points under vacuum and rotary evaporation versus simple distillation.
A complete rotary evaporation station has four parts. Each should be planned together:
| Component | Role | ALab™ example |
|---|---|---|
| Rotary evaporator | Rotates the flask in a heated bath and condenses vapor | RE100-Pro: 50–3000 mL flasks; 20–280 rpm; bath RT to 180 °C; motorized 150 mm lift |
| Vacuum source | Lowers the boiling point | C410 diaphragm vacuum pump: oil-free; 13 mbar ultimate vacuum; 25 L/min free air |
| Vacuum controller | Holds a set pressure to limit bumping and improve recovery | VC100 vacuum controller: 1–1000 mbar; 5 programs × 5 steps |
| Condenser cooling | Removes heat from the condenser | DCP-5 recirculating chiller: −20 to 40 °C; ±0.3 °C; 700 W cooling capacity at 20 °C |
Where bench space or a chiller loop is a constraint, the RE100-CV combines a self-cooling condenser (room temperature to −40 °C) with a built-in vacuum controller (1–1000 mbar). The rotary evaporator category page lists the rest of the range. Flammable solvent work also triggers ventilation and fire-protection requirements; see NFPA 45 and your institution's chemical hygiene plan [14][1].
5. Measurement: UV-Vis and Electrochemistry Meters
Measurement instruments turn a sample into data. Two classes appear on most research lab equipment lists.
UV-Vis spectrophotometers
UV-Vis spectrophotometry is used to determine concentrations, follow reaction kinetics, check nanoparticle formation and assess nucleic acid or protein samples. Quantitative work relies on the Beer–Lambert law, which relates absorbance to concentration and optical path length [9]. Specifications that matter:
- Wavelength range. Visible-range instruments typically start around 320 nm. Work in the ultraviolet (for example, nucleic acids at 260 nm) needs a UV-capable instrument with quartz cuvettes.
- Spectral bandwidth. Narrower bandwidth resolves sharp absorption features more faithfully.
- Single or double beam. A double-beam design compares sample and reference signals during a measurement, which can help compensate for changes in the light source and improve measurement stability during long scans or kinetic measurements.
The SP-1000/1100 series includes four single-beam models (SP-V1000, SP-UV1000, SP-V1100 and SP-UV1100); the SP-UV1100 covers 190–1100 nm, and bandwidths of 4.0 nm or 2.0 nm are available. The double-beam SP-UV3101 covers 190–1100 nm with 1.0 nm bandwidth and a 10.1-inch touchscreen. See the spectrophotometers page.
pH, conductivity and dissolved oxygen meters
pH is the most common measurement in chemistry and biology alike, and buffer preparation alone justifies a dedicated meter. The IUPAC recommendations on pH measurement define the procedures and standard buffers that make pH values traceable [8]. Conductivity is used to check water purity and electrolyte concentration (ISO 7888 describes a standard water-quality method [11]). Dissolved oxygen (DO) matters in environmental, electrochemical and fermentation work (ISO 17289 covers the optical sensor method [10]).
Choose the form factor by where the measurement happens:
| Form factor | Best for | ALab™ example |
|---|---|---|
| Benchtop meter | Daily buffer preparation, multi-point calibration, data logging at a fixed station | Benchtop Meters, Classical Series: seven pH/conductivity/DO models; 7-inch display; automatic calibration; automatic temperature compensation |
| Portable meter | Measurements at several benches, in a fume hood or in the field | Portable Meters: eight models; 3.5-inch TFT display; IP57; for example the DPCD10 for pH, conductivity and DO |
| Pocket tester | Quick checks, teaching labs, spot measurements | Pocket Testers: eight models; IP67; replaceable probes |
IP ratings are defined in IEC 60529 and describe protection against solids and water ingress [13]. All meters are listed on the electrochemistry meters page.
6. Temperature Control: Dry Baths, Thermo Mixers and Chillers
Many steps need a sample held at a set temperature: enzyme incubation, denaturation, sample storage during handling, or cooling a condenser. The right tool depends on format and range.
- Dry baths hold tubes in aluminum blocks with no water, which avoids contamination and evaporation issues. The H100-Pro dry bath covers room temperature +5 °C to 100 °C with ±0.5 °C accuracy, uses magnetic heat blocks and supports six-step programs. The low-profile HB120-S / HB120-SC / HB60-S series offers fixed-temperature and adjustable options up to 120 °C.
- Thermo mixers combine temperature control with orbital mixing, useful for enzymatic reactions and lysis. The HCM100-Pro / HM100-Pro mixes at 200–1500 rpm with a 3 mm orbit. For the HCM100-Pro's cooling range, consult the live product page.
- Cooling blocks replace ice buckets for keeping enzymes and samples cold during setup. The CT4-Pro cooling thermo block holds 4 °C with ±0.5 °C accuracy and reaches 4 °C in 10 minutes or less.
- Recirculating chillers cool condensers, reactor jackets and other equipment. The DCP-5 covers −20 to 40 °C with ±0.3 °C stability and a 4.5 L tank; its published cooling capacity is 700 W at 20 °C. Cooling capacity falls as the setpoint drops, so check the manufacturer's figures for your working temperature rather than estimating them.
The dry baths and thermal control page lists the full range.
7. The Molecular Biology Bench
Molecular biology labs build on the same liquid handling, centrifugation and temperature-control core, then add instruments for nucleic acid work. Everything in this section is described for research use only. Biosafety requirements depend on the organisms and materials involved and should follow a protocol-specific risk assessment as described in the BMBL [3].
Thermal cyclers and real-time PCR
A thermal cycler is the central instrument of a PCR workflow. Compare block format (96-well is standard), ramp rate, temperature range and whether a gradient function is available for optimizing annealing temperatures. The TC1000-G-Pro and TC1000-S+ are 96-well Peltier cyclers with a 0–105 °C block range and up to 5 °C/s ramping; the TC1000-G-Pro adds a 30–105 °C gradient. For quantitative PCR, the Accurate 96 real-time PCR system offers 4 or 6 fluorescence channels and up to 6.0 °C/s ramping. When reporting qPCR results, the MIQE guidelines describe the minimum information that should accompany published experiments [17]. See the PCR thermal cyclers page.
Electrophoresis and gel imaging
Agarose gel electrophoresis separates DNA fragments by size and remains a routine check after PCR and digestion [18]. A basic setup includes a tank, a power supply and a way to visualize the gel.
| Item | ALab™ example | Published figures |
|---|---|---|
| Horizontal (agarose) tank | DL-SUB01 | 60 × 75 mm gel; 300 mL buffer |
| Vertical (polyacrylamide) tank | DL-Mini01 | Two 8.3 × 7.3 cm gels |
| Wet transfer tank | DL-ZY02 | 110 × 90 mm transfer area |
| Power supply | DEP-300 / DEP-600 / DEP-300HC | Constant voltage, current or power; 0–300 V or 0–600 V |
| Gel visualization | TI470 blue light transilluminator | 470 nm blue LED with amber filter, for compatible nucleic acid stains |
Blue-light transillumination with a compatible stain avoids working over a UV source. UV exposure limits for skin and eyes are set out in ICNIRP guidelines [19]; if a lab does use UV transilluminators, plan face and eye protection accordingly. See the electrophoresis and gel imaging and transilluminators pages.
Nucleic acid extraction
Labs that process many samples may automate extraction. The MNE-96-03 automated nucleic acid extraction system uses magnetic beads to process 1–96 samples at 50–1000 µL. Smaller labs often begin with manual column or bead kits and a microcentrifuge, and add automation once sample numbers justify it.
8. Lab Equipment Checklists by Lab Type
The four checklists below are editorial suggestions from ACS Material to help you plan. They are not standards, regulations or a complete inventory. Your institution's safety office, chemical hygiene plan and biosafety review determine what is required [1][3]. Priorities:
- Essential: needed from the first day of routine work.
- Common: found in most labs of this type; add within the first months.
- Can wait: useful once throughput or methods grow; often shared at first.
Items without a link are general lab infrastructure or categories outside this guide.
Materials chemistry lab (synthesis of nanomaterials, coatings, functional materials)
Much materials science lab equipment overlaps with the synthetic chemistry list; the difference is a stronger emphasis on dispersion, particle collection and optical characterization.
| Item | Priority | Why it is on the list | Example |
|---|---|---|---|
| Fume hood or other local exhaust | Essential | Solvent and precursor handling | Facility infrastructure |
| Analytical balance | Essential | Weighing precursors | — |
| Single-channel pipette set (0.5–1000 µL) plus 1–10 mL | Essential | Precursor and reagent transfer | Toppette |
| Bottle-top dispenser for solvents | Common | Repeated solvent dispensing | DispensMate-Pro |
| Magnetic hotplate stirrer with external sensor | Essential | Heated synthesis at controlled solution temperature | MS7-H550-Pro |
| Multi-position stirrer | Can wait | Parallel screening | MS-H-S10 |
| High-shear homogenizer | Common | Dispersing particles, emulsions | D-500 |
| Overhead stirrer | Can wait | Viscous slurries and polymer solutions | OS-Pro series |
| High-speed centrifuge | Essential | Collecting and washing particles | DG1616 |
| Vortex mixer | Common | Redispersing pellets | MX-S |
| Rotary evaporator with vacuum pump | Common | Solvent removal and recovery | RE100-Pro + C410 |
| UV-Vis spectrophotometer | Common | Optical characterization, concentration | SP-UV3101 |
| pH meter | Essential | Precursor solutions and pH-controlled growth | Benchtop Meters |
| Drying oven and furnace | Common | Drying and annealing | — |
Synthetic chemistry lab (organic and inorganic synthesis)
| Item | Priority | Why it is on the list | Example |
|---|---|---|---|
| Fume hoods sized for the number of chemists | Essential | Primary engineering control for volatile chemicals | Facility infrastructure |
| Analytical and top-loading balances | Essential | Weighing reagents and products | — |
| Magnetic hotplate stirrers, one per chemist | Essential | Most reactions are stirred and heated | MS-H-Pro+ |
| Heating mantles with stirring for round-bottom flasks | Common | Reflux with even heating | MS-H280-B / F series |
| Rotary evaporator | Essential | Workup and solvent removal | RE100-Pro |
| Oil-free vacuum pump | Essential | Rotary evaporation, filtration | C410 |
| Vacuum controller | Common | Reproducible evaporation, less bumping | VC100 |
| Recirculating chiller | Common | Condenser cooling without tap water | DCP-5 |
| Bottle-top dispensers for common solvents | Common | Safer repeated dispensing | dFlow |
| Electronic burette | Can wait | Titrations and standardization | dTrite |
| Overhead stirrer | Can wait | Scale-up, viscous mixtures | OS-Pro series |
| Benchtop centrifuge | Common | Separating precipitates | D3024 |
| Flammables storage cabinet | Essential | Solvent storage per fire code | — |
| Glassware, Schlenk line or inert gas manifold | Common | Air-sensitive chemistry | — |
Molecular biology lab (research use only)
| Item | Priority | Why it is on the list | Example |
|---|---|---|---|
| Biosafety cabinet, as required by risk assessment | Essential | Containment for biological materials | Facility infrastructure |
| Single- and multichannel pipettes, autoclavable | Essential | All reaction setup | MicroPette Plus |
| Electronic pipette | Common | Long plate setups, repeat dispensing | dPette+ |
| Pipette controller | Essential | Serological pipettes for media and buffers | Levo Plus |
| Refrigerated microcentrifuge | Essential | Extractions and cold spins | D1524R |
| Palm spin-down centrifuge | Common | Collecting liquid in tubes and strips | D1008 |
| Vortex mixer | Essential | Mixing and resuspension | MX-S |
| Dry bath | Essential | Incubation and denaturation | H100-Pro |
| Thermo mixer | Common | Heated mixing for enzymatic steps | HCM100-Pro |
| Cooling block | Common | Keeping enzymes cold during setup | CT4-Pro |
| Thermal cycler | Essential | PCR | TC1000 series |
| Electrophoresis tank and power supply | Essential | Checking PCR products and digests | DL-SUB01 + DEP series |
| Gel visualization | Essential | Reading gels | TI470 |
| Real-time PCR system | Can wait | Quantitative assays | Accurate 96 |
| Aspiration system for liquid waste | Common | Safer liquid waste collection | SafeVac |
Teaching lab (undergraduate chemistry and general science)
| Item | Priority | Why it is on the list | Example |
|---|---|---|---|
| Eyewash and safety shower within reach | Essential | Emergency response | Facility infrastructure |
| Fume hoods for demonstration and volatile work | Essential | Exposure control | Facility infrastructure |
| Mechanical pipettes, one set per bench group | Essential | Measurement technique training | Toppette |
| Bottle-top dispensers at the reagent station | Common | Faster, cleaner reagent distribution | DispensMate-Pro |
| Magnetic stirrers (non-heating) | Essential | Dissolution, titration | MS-S |
| Magnetic hotplate stirrers with overheat protection | Essential | Heated experiments | MS-H-Pro+ |
| Pocket pH testers | Essential | One per student pair | Pocket Testers |
| Benchtop pH/conductivity meter | Common | Instructor reference, buffer preparation | Benchtop Meters |
| Visible spectrophotometer | Common | Beer–Lambert experiments | SP-1000/1100 series |
| Electronic burette | Can wait | Titration demonstrations | dTrite |
| Vortex mixer | Common | Mixing small samples | MX-S |
| Small benchtop centrifuge | Can wait | Precipitate separation exercises | D3024 |
| Top-loading balances | Essential | Weighing | — |
9. Shared Utilities: What Sets the Grade of Equipment You Need
Utilities and building conditions often decide which version of an instrument a lab can run. Check these before finalizing the list.
Bench space and layout
Measure usable bench depth and width, not just the instrument footprint. Rotary evaporators need clearance for the lift and the flask; centrifuges need clearance around the lid and a stable surface; hotplates need distance from flammable materials and from instruments sensitive to heat. Place the instruments in the order the sample travels, so samples move along the bench rather than back and forth across the room.
Electrical supply
Heating instruments, centrifuges and chillers draw the most power. Check the rating published on each product page, for example 550 W for the MS-H-Pro+, 1050 W for the MS7-H550-Pro or 500 W for the D-500. Then ask your facilities team how circuits are distributed along each bench. Avoid daisy-chained power strips for heating equipment. Laboratory electrical equipment is designed and tested against IEC 61010-1 [12]; confirm voltage and plug type for your region before ordering.
Ventilation
Ventilation is a building system, not a benchtop purchase, but it shapes the equipment list. The number of fume hoods limits how many chemists can run volatile reactions or rotary evaporation at once. Laboratory ventilation is addressed in ANSI/ASSP Z9.5, fume hood performance testing in ANSI/ASHRAE 110 and fire protection for chemical laboratories in NFPA 45 [15][16][14]. Under the OSHA Laboratory Standard, the chemical hygiene plan must address the proper function of fume hoods [1].
Chemical resistance
Check what your reagents touch. Wetted parts on dispensers and vacuum components, plate coatings on hotplates and the housing of benchtop instruments all have chemical-resistance limits. For example, the VC100 lists PTFE, PP, silicone and ceramics as its vapor-contact materials, and the DispensMate-HF uses ceramic and FEP for hydrofluoric acid. When a chemical is not listed, ask before you buy.
Water, emergency equipment and waste
Plan purified water supply, eyewash and safety shower locations (ANSI/ISEA Z358.1 is the common US reference [21]), and separate waste streams for solvents, aqueous waste and biological waste. These decisions affect where equipment can go.
10. Frequently Asked Questions
What basic laboratory equipment does every new research lab need?
Most new labs need calibrated pipettes, a magnetic hotplate stirrer, a vortex mixer, a centrifuge suited to their tube formats, a pH meter and an analytical balance, plus the fume hood and safety equipment set by their chemical hygiene plan. The rest depends on the lab type; see the checklists in Section 8.
How should I prioritize a new lab setup checklist?
Rank equipment by how often a missing or shared instrument stops daily work. Items everyone uses every day come first. Items used for one technique occasionally can be shared or added later. Our checklists mark each item Essential, Common or Can wait as an editorial starting point.
What is the difference between a chemistry lab equipment list and a molecular biology list?
Both share liquid handling, mixing, centrifugation and temperature control. Chemistry labs add fume hoods, rotary evaporation, vacuum pumps and higher-temperature heating. Molecular biology labs add autoclavable pipettes, refrigerated microcentrifuges, thermal cyclers, electrophoresis and gel imaging, and follow a biosafety risk assessment.
Should I specify centrifuges by rpm or by ×g?
By ×g (relative centrifugal force). RCF depends on both speed and rotor radius, so rpm alone does not describe the force on the sample. Write protocols in ×g and confirm that the instrument and rotor reach it.
Do I need a rotary evaporator in a new lab?
If your lab regularly removes solvent from heat-sensitive samples or recovers solvent, a rotary evaporator may be useful. Whether it is appropriate depends on the solvent, sample sensitivity, batch size, throughput and available alternatives. Our rotary evaporation guide explains how to size and set up a station.
Are these checklists a safety standard?
No. They are editorial suggestions to help with planning. Requirements come from regulations, your institution's chemical hygiene plan and biosafety review, and the manufacturer's instructions for each instrument.
11. Keep Exploring the Lab Essentials Hub
- Rotary Evaporation: The Complete Guide: principles, setup and solvent recovery.
- How a Rotary Evaporator Works: Five Steps
- Rotary Evaporation vs Simple Distillation
- Solvent Boiling Points Under Vacuum
- Why the Flask Rotates: Films, Area and Bumping
Browse all ALab™ Lab Essentials.